Download Heat of Solution and Voltmeter Circuits - Experiment 4 | CHEM 461 and more Lab Reports Physical Chemistry in PDF only on Docsity! Revised 6/06
Chemistry 461
EXPERIMENT 4—HEAT OF SOLUTION AND VOLTMETER CIRCUITS
1, WASTE DISPOSAL
1. Ail salt solutions (including BaSOq precipitate)—Waste bottle labeled
“Experiment 4 Salt Solutions,” located in the hood. Clean up spills in the hood
promptly.
DO NOT FILL THE WASTE JUGS COMPLETELY TO THE TOP!
2. Reaction products from the standardization of the Parr Solution Calorimeter and
any remaining HCl—Adjust pH to between 5.5 and 12 using NaHCO; (check
final pH with pH paper), then wash down the drain.
3. Any remaining solid salts (including TRIS)—Waste jar labeled “Experiment 4
Waste Salts,” located on the shelf under the windows on the north wall. Clean up
spills in the balances promptly.
Ii. INTRODUCTION
In this experiment, you wiil make experimental measurements of the heats of
solutions for the following reactions:
Dissolution Reaction No. of Runs
NaNO; in H20 3
Ba(NO3)2 in H2O 2
Na,SOx in H20 2
Ba(NO3)2 + NagSO, in H,O 2
Ba(NO3)o in 0.2 M NaSOx 1 (Extra Credit)
The number of experimental runs required is also listed above.
The first three measurements directly give the heats for the following three reactions:
I Ba(NO3), (s) + HyO ——> Ba** (aq) + 2 NO} (aq) AH,
IL Na,SO, (s} + H,O ——> 2Na” (aq) + so, {aq) AHy
Il. NaNO, (s) + H,O ——> Na* (aq) + NO; (aq) AB
The reaction for the solvation of BaSO4 may be written:
IV. BaSO,(s) + H,O ——> Ba” (aq) + SO; (aa) AByy
The actual fourth reaction you run in the laboratory is
4, Na,8O,(s) + Ba(NO3), (s) + H,O —>
BaSO, (8) + 2.Na* (ag) +2. NO (aq) AH,,
which involves formation of a precipitate, and is the sum of reaction I plus reaction II
minus reaction IV. Hence we have that
AHy = AH, + AH, - AH.
Finally, note the possibility of the solid phase reaction:
V. Ba(NO,), (s) + Na,SO,(s) —>
BaSO, (s) + 2 NaNO; (s) AHy,
which can be written approximately as
1+0-20 -IV
so
AH, = AH, + AHy - 2AHg - AHy.
Note that AHy may be calculated from standard heats of formation, easily found in the
Handbook of Chemistry and Physics. This serves as a literature check on the heats
determined above. Note also that literature values for some of the heats of solution are
posted on the bulletin board in the lab.
The final reaction [Ba(NO3)2 in 0.2 M Na2SO,] can be run for extra credit. It is
simply a linear combination of two of the other measured reactions. It thus serves as an
internal check on the measured heats.
These runs can be made on either a “homemade” calorimeter, or on the Parr 1455
Solution Calorimeter. You are, however, required to make no less than 2 and no more
than 4 of these runs on the Parr instrument. The rest must be made on a “homemade”
calorimeter. Since the “homemade” calorimeter has poorer insulation and other
problems, higher concentrations are always used when using it.
IV. GENERAL USE OF “HOMEMADE” APPARATUS
The apparatus is shown schematically in Figure 1. It consists of a thermally
insulated Dewar flask for aqueous solutions. A thermistor (temperature-sensitive
resistor) is placed into the water in this flask to measure its temperature. This is
accomplished by measuring the thermistor resistance (R,) via a Wheatstone bridge
circuit, which includes a galvanometer, G. Such measurements are well described
in the “electricity” section of your undergraduate physics textbook. It is also
necessary to read in such a book about the temperature coefficient of resistivity to
learn how a thermistor works. Also placed into the flask of water is an electrical
heater. The heats of solution will be counterbalanced or mimicked exactly by the
energy dissipated by this electrical heater. Your physics textbook will also tell
you how to calculate this electrical energy by knowing the time (t) over which a
fixed current (i) flows through this heater (resistor, Ry) and the voltage drop (V)
that is maintained across this heater while this current flows. What is this
necessary mathematical relationship? Voltages are measured using a
potentiometer (digital voltmeter), current is determined by measuring the voltage
drop across a 0.100 ohm calibrated resistor in series with the heater.
Special Funnel Thermistor Leads
(Red)
Heater Leads Thennistor
Black/White)
[—— Dewar Flask
Heating Coil
Magnetic Stir Bar
Stirring Motor
Figure ta. “Homemade” Calorimeier for Heat of Solution
Digital
Voltmeter
C
L isv
T Dry Cell
Ammeter 1 Coil I Wheatstone
Bridge
Figure 1b. Circuit Diagram of “Homemade Calorimeter”
General Measurement Instructions:
The following preliminary operations should not be done at this time. They are
merely operations that you will need to repeat several times in the Experimental
Procedure section.
Voltage Measurements:
Plug the leads from the calorimeter console labeled “POT” into the digital
voltmeter (DVM), the black lead into the jack labeled “COM” and the red lead
into the lead labeled “V-Q”. To conserve the battery, leave the DVM power off
except when actually making a reading. Whenever turning “on” a DVM to read
voltages or currents, always set it to the least sensitive (maximum range) before
turning it “on”, then adjust for maximum precision. Unless otherwise noted, set
the toggle switch on the calorimeter console so that actual voltages are read (i.e.,
in the “Vx1” position). When in the “V/100” position, the DVM sees 1/100th of
the true voltage. (Prove in your lab report why this is true.)
Adjustment of Heater Current to 1.000 Ampere:
Caution: Never turn on heater current unless the heater coil is immersed in
water.
a, Place the knife switch on the front of the plexiglass case in the CURRENT
position. This puts the DVM leads across the 0.100 Q resistor. (See Figure
1b.)
b. Switch “on” the power to the DVM and the function and range dial to “DCV,
200m.” On this scale, the maximum voltage you can read is 200 mV.
c. Bring the voltage reading to exactly 100.0 mV using the current-adjust control
(the multi-turn dial on the front of the plexiglass case). Note that the analog
(needle-type) ammeter may not read 1 ampere after the current has been
adjusted. Do not be concerned about this. The value that is determined by
using the DVM is much more accurate; in this case the analog ammeter should
be used only as an on-off indicator rather than a measuring device.
d. Turn off the power to the DVM.
What you have done is to adjust the voltage across a calibrated 0.100 ohm resistor
{in series with the heater coil) to be exactly 0.1000 volts (100.0 mv). Thus, from
Ohm’s Law, the current flowing through the resistor (and the heater coil) is 1.000
ampere.
Determination of the Voltage Drop Across the Heater Coil:
a. Place the knife switch in the VOLTAGE position. As shown in Figure 1b, this
puts the DVM leads across the heater resistance, Ry.
b, Measure the voltage drop across Ry (i.e., the heater coil) using the DVM.
c. Switch the toggle switch temporarily to the “V/100” position, and repeat the
measurement of step (b) above. Sometimes, the actual voltage drop across
some element in a circuit is too great to be measured directly using a normal
DVM. (Nowadays this usually happens only above a few kV.) What you
have done is to measure the potential drop across the smaller resistance of a
potential divider in parallel with Ry. The potential divider is set in a ratio of
100:1; this means that the actual voltage drop across the heater should be 100
times that value read on the DVM.
Note: When determining this voltage, you may observe some fluctuations in
the reading; at other times it may be quite stable. If these fluctuations, which
are the result of fluctuations in the AC line voltage, are present, use the best
average value for the voltage that you are able to obtain.
d. Switch the toggle switch back to the “Vx1” position and turn off the DVM.
8.
10.
1.
As the water temperature nears 25°C, turn off the heater; note that the
temperature continues to rise for several seconds. Try to determine the
approximate magnitude of this temperature lag so that you will not overshoot the
required temperature. (This may be easier to determine by monitoring the change
in the thermistor’s resistance; see below.) Approach the 25° mark using the
“momentary” switch. When the temperature has been adjusted to exactly 25°,
determine the resistance of the thermistor which corresponds to this temperature.
This exact resistance value should serve as the reference point for all
experimental measurements, since it is a more accurate measurement to
reproduce this reference temperature than to measure the temperature with a
thermometer.
To estimate the magnitude of the temperature lag of the heater by monitoring the
thermistor resistance, adjust the decade resistors to bring the galvanometer on
scale (to the left of zero) at some temperature below 25°. (The heater should be
on.) Turn the heater off and observe how long the galvanometer continues to
rise, and measure the magnitude of this rise.
When the water temperature has been adjusted to exactly 25° and the thermistor
resistance measured, REMOVE THE LONG-STEM DIGITAL
THERMOMETER and insert the “special” sample addition funnel.
Set the electric timer to zero time; make sure the stirrer is operating.
Note: In all runs, the total stirring time must be determined; this is the elapsed
time between the addition of the salt and the final temperature adjustment of the
calorimeter contents. This will be required to calculate the “stirring correction”
(step 16 below).
You are now ready to make your first practice run measurement of the heat of
solution of NaNO3. Add the first sample of NaNO3 to the water, and start the
digital stopwatch. Close the telegraph key; the galvanometer in the thermistor
circuit should deflect to the left indicating a drop in temperature (thermistor
resistance), which occurs due to the endothermic reaction:
NaNO,(s) + H0(é) ——> Na” (aq) +NO3(aq)
Tum on the heater to counteract the cooling effect of the salt and check the
approach to 25° by monitoring the thermistor resistance. As the endpoint of 25°
is approached, change to the “momentary” heater switch so as to insure that the
endpoint is just reached, but not exceeded. The apparatus is not equipped with a
calibrated cooling device so no correction can be made for overheating; any run
in which the endpoint is overshot must be discarded and repeated. When the
endpoint is reached, stop the digital stopwatch and record the elapsed time (the
total stirring time). Record the heating time required (from the analog timer).
The energy (heat) of solution can be determined by multiplying this heating time
10
12.
13.
14,
15.
16.
by the heater power. Do an energy balance to prove this is the case before the lab
starts.
Check the heater current and voltage drop across the heater after each run. If
there has been any change in either value, the average must be used in your
calculation of the heater power (P=I-V). Remember that the heater must be
ON before measuring these quantities.
Repeat the above procedure for the two real samples of NaNO; and for the 2
samples of Ba(NO3)2.
For the (slightly) exothermic salt Na»SO, a different procedure must be used.
The water temperature is adjusted to the same reference point (thermistor value)
as for the other salts, and the Na,SO, is added through the funnel. However, in
this case, the heater is not turned on. Instead, the change in thermistor resistance
(temperature) is monitored. The galvanometer needle should deflect to the right
of zero upon salt addition until it has reached a stable value indicating that all of
the Na2SOy has been dissolved. When this occurs, the new thermistor resistance
(temperature) is measured by adjusting the decade resistors to bring the
galvanometer back to zero; this value is recorded. Reset the decade resistors
back to that value corresponding to 25° and cool the solution below this point by
immersing a cold glass rod. Turn on the heater and bring the temperature of the
solution to exactly the 25° thermistor reading. Reset the decade resistors to that
“temperature” value which was obtained after the dissolution of the Na)SOq, and
reset the timer to zero. Turn on the heater and record the time required to heat
the solution to this final “temperature;” this time is treated in the calculations
exactly as the times recorded for the endothermic salts. Prove by energy balance
that this is the case.
After both runs have been made on each of the individual salts, the sample
consisting of a mixture of Ba(NO3)2 and NazSO, is run. The overall process of
solution plus reaction will still be endothermic in this case.
After the last heat of solution run is made, the energy contribution to the system
by stirring plus that energy lost due to non-perfect insulation (the so-called
“stirring correction”) is determined. Adjust the initial water temperature to 25°
as in the previous runs, Run the stirrer for about 30 minutes (recording the actual
length of time the stirrer operates); at the end of this time, determine whether the
temperature has increased or decreased. If the temperature has increased, the
same procedure as for an exothermic salt is followed. Measure the new
thermistor resistance, cool the water to below 25° with a cold glass rod, adjust
the water temperature to 25° with the heater, and determine the length of time the
heater must be on in order to raise the water temperature to that which was
obtained after 30 minutes of stirring. From this time and the heater current and
voltage, the energy input per second of stirring is calculated. If the water
temperature has decreased after 30 minutes (indicating the system has lost
1
17.
18.
energy), the procedure for an endothermic salt is followed. Determine the length
of time the heater musi be on in order to heat the water back to 25°. From this
time, the energy loss per second is calculated. This value should be used as a
correction in your energy balance equations.
At the end of each day be sure to do the following:
a) Turn HEATER AND TIMER and MAIN POWER switches (on the
calorimeter console) to OFF.
b) Turn stirring motor off.
c) Turn Fluke Electronic Galvanometer and DVM function switches to OFF.
d) Unplug the leads to the DVM and return it to the stockroom.
When you have finished the experiment, don't forget to remove any unused
samples from the drying oven. Correctly dispose of any salts you have stored in a
desiccator, and return the desiccator to the shelf.
12
Vii. EXPERIMENTAL PROCEDURE FOR THE PARR CALORIMETER
CAUTION! The glass sample cell must only be handled by either the large diameter
glass base or the plastic collar. The small diameter glass stem is thin-walled and fragile
(and the cell is expensive) and will break very easily if any stress is placed on it.
A. Calibration:
1. Tum on the calorimeter power (switch on rear of instrument on upper left) and the
chart recorder; allow them to warm up for at least 10 minutes before making a run.
Check that the recorder CM/MIN-OFF-CM/HR switch is set to OFF, the
STANDBY/RECORD switch is set to RECORD, and that the INPUT RANGE
SELECTOR is set to 1 volt. (See Figure 3 for the location of recorder controls.)
Raise the pen from the paper using the PEN LIFT LEVER and remove the cap from
the pen; place the cap in a location where it may be easily found at the end of the lab
period. Rotate the ATTENUATOR control to its counter—clockwise limit.
In order to make good comparisons with the numbers obtained with the “homemade”
calorimeter, all runs should be made at 25°C. Before starting the runs, place an
adequate supply of the solvents used, either 0.1 M HC! or deionized water, in one of
the 25°C water baths to equilibrate; ten or fifteen minutes should suffice.
The first operation is the calibration of the calorimeter, i.e. determination of the
calorimeter's heat capacity. The calorimeter should be calibrated each day, or
whenever operating conditions change (e.g., if a new Dewar flask or sample cell is
used). The calibration should be repeated at least once, more if the results do not
closely agree, It is essential that the calorimeter be accurately calibrated because all
subsequent results will depend upon the calibration.
The calibration consists of reacting a known quantity of
tris(hydroxymethyl)aminomethane, commonly call TRIS, with dilute hydrochloric
acid. TRIS is an organic base for which the heat of reaction with hydrochloric acid is
accurately known; its formula is (HOCH2)3;CNHz. The TRIS sample should be
ground before the calibration runs, but need not be dried.
In each of the runs it will be necessary to set the calorimeter span (number of degrees
Celsius represented by a full-scale deflection of the recorder) and offset (the
temperature corresponding to a recorder reading of zero). These are most easily
estimated by first performing a trial run.
2. Remove the drive belt from the pulleys and remove the cover from the calorimeter.
Lift the Dewar flask out of the metal can; dry the Dewar thoroughly if necessary.
Tare the Dewar on the large triple beam balance and weigh about 100g of
(approximately) 0.1 M HCl to the nearest 0.1g. Place the black plastic spacer ring
around the Dewar and place the filled Dewar in the calorimeter.
15
PEN COVER:
rtf) J
ATIENTUATOR
PEN LIFI/PAPER
RELEASE LEVER STAND BY/RECORD SWITCH | INPUT RANGE
* SELECTOR
ZERO CONTROL
PER SPROCKETS CHART SPEED
PA oe — +: SELECTOR
* < ec
UW ® iil fl e £
PAPER HOLD DOWN POWER INDICATOR CM/MIN-OFF-CM/HR
CORE SUPPORT KNOB CORE SUPPORT DISCS
CHART PAPER
CONNECTOR
col
oo
ON-OFF POWER ENTRY MODULE CHASSIS LOCAL/REMOTE
SWITCH Wit VOLTAGE SELECT GND SWITCH
AND FUSE
SIGNAL-
SIGNAL+
Figure 3. Top and Rear Panels of Chart Recorder
16
. Check that the glass sample cell, the glass push rod, and the Teflon sample dish are all
clean and dry. Remember to handle the sample cell only by the plastic collar or the
large diameter glass base. Weigh ~0.5g of TRIS to the nearest 0.1 mg directly into
the tared Teflon sample dish; be careful not to drop any of the sample into the push
rod socket in the center of the dish or on the lip around the outside of the dish. Given
that the molecular weight of TRIS is 121.14, verify that the amount of 0.1 M HCl in
the Dewar is sufficient to completely neutralize your TRIS sample. Set the dish on a
flat surface and carefully press the glass cell over the dish to seal the sample inside.
Be careful to grasp only the sides of the thick-walled giass cell during this operation,
not the stem or collar; press the glass cell down over the Teflon dish as far as it will
go to insure a good seal. Attach the cell to the calorimeter cover stirring shaft by
sliding the plastic collar of the cell onto the shaft as far as it will go and turning the
thumb screw finger tight.
Hold the cover in a horizontal position and lower it carefully until the bottom of the
cell rests on the bench top; then, while firmly grasping the large diameter glass cell
base, insert the glass push rod through the pulley hub and press the end of the rod into
the socket of the Teflon sample dish.
. Lower the cover assembly with the cell into the Dewar and set the cover in place on
the calorimeter can. Rotate the cover so that the thermistor probe opening is pointed
to the rear of the calorimeter; install the thermistor probe in the cover opening and
press the bushing firmly into place to anchor the probe in its proper position. Check
that the two plugs in the cover are completely seated; they can occasionally be
knocked loose if the cover rests on a flat surface.
Note: Handle the thermistor probe carefully. It can be damaged if dropped.
. Drop the geared drive belt over the motor and stirrer pulleys. The belt may appear to
be unusually loose, but it is intended to operate under very light tension to minimize
friction in the stirrer bearing. Turn on the stirring motor by pressing (FI). Allow the
system to equilibrate for several minutes before setting any calorimeter parameters.
, Set the recorder CHART SPEED SELECTOR to a setting of 3. (Runs will be
recorded at a chart speed of 3 cm/min.) Start the chart paper by setting the CM/MIN—
OFF-—CM/HR switch to CM/MIN, and lower the pen using the pen-lifting lever on the
left side of the recorder.
Note: If the sample is slow to react or dissolve, it may be desirable to use a chart
speed of 2 cm/min or even 1 cm/min. This can also be determined during the trial
run.
. Observe the liquid crystal display above the calorimeter keyboard. The top line gives
prompts when entering parameters such as the span and offset. The bottom line gives
various output data, the majority of which is used only for other applications.
Observe the bottom line until you see “CHNL 1” followed by a temperature reading;
17
12.
13.
14.
15,
16.
17.
18.
When a linear preperiod drift line has been established, combine the reactants by
pressing the glass push rod downward to drop the sample out of the rotating cell. Use
the palm of the hand to push straight down on the rod; do not grasp the rod, as this
will interrupt the rotation of the cell and the stirring of the solution. Very soon after
combining the reactants, there should be a pen deflection corresponding to the
temperature change within the calorimeter.
Allow the stirrer to operate during the postperiod until the drift line becomes linear
and remains linear for at least 3 minutes indicating that the reaction has reached
completion. Usually this condition will be reached within 10 minutes or less after
combining the reactants.
After a satisfactory postperiod drift line has been established, set the CM/MIN-OFF—
CM/HR switch to OFF and raise the pen from the chart. Note that raising the pen
disengages the chart drive so that the thermogram may be easily pulled out and torn
off against the paper hold down bar. Do not try to manually advance the chart unless
the pen is raised! Label the thermogram, identifying the reactants; record all pertinent
parameters used during the run.
Stop the calorimeter stirring motor:
Press
Press
Note: When using the (SHIFT) key, it need not be held down as on a keyboard; simply
press it once and then press the second key.
Remove the drive belt.
Remove the thermistor probe and wipe dry.
Carefully raise the calorimeter cover and lift the cell assembly out of the Dewar flask.
Remove the sample dish from the end of the push rod, then remove the rod; carefully
loosen the thumbscrew holding the glass sample cell to the stirring shaft and gently
remove the sample cell. Remember: Handle the glass sample cell only by the large
diameter base or the plastic collar, not the small diameter stem. Lift the Dewar flask
out of the calorimeter can and empty it. Wash and dry all wetted parts carefully.
At the end of the day: Carefully wash and dry all wetted parts of the cell assembly
and the Dewar flask. Place the Dewar in the calorimeter can; replace the cover and
drive belt. Check that the recorder pen is lifted from the chart, replace the pen cover,
and turn off the calorimeter and recorder.
20
Calculation of the Parr calorimeter constant:
In order to determine the net temperature change produced by the reaction, it is
necessary to locate a point on the thermogram at which the temperature reached 63%
of its total rise (or fall). This can be easily done using the graphic method described
below and shown in Figure 4.
1. Place a straight edge over the preperiod drift line and extend this line well past the
point at which the cell was opened to start the reaction.
2. Move the straight edge to the postperiod drift line and extrapolate this line backward
to the time when the cell was opened. If there are fluctuations in the drift lines due to
noise or other variations in the signal, use the best average when drawing the
extrapolations.
3. Using a centimeter scale, measure the vertical distance, R, between the two
extrapolated lines at a point near the middle of the reaction period, and convert it to
temperature units using the chart recorder’s scale factor from above. This
corresponds to the temperature change due to the process, AT,. As shown in Figure 4,
this is most accurately measured at a point where this temperature change is ~63%
completed, since this minimizes errors associated with differing slopes in the
preperiod versus postperiod.
Note that while Figure 4 illustrates an exothermic reaction, the same procedure is
used for an endothermic reaction.
4. The energy change, Q, is calculated by
Q=AT, «©
where e is the energy equivalent of the calorimeter and its contents:
© = Soy + (mo sow)
where Ccqj = the calorimeter constant (or heat capacity of the empty calorimeter),
Msoly = the mass of the solvent, and
Csoly = the specific heat of the solvent.
The calorimeter constant is then easily calculated given that 58.738 calories per gram
of TRIS is evolved when TRIS is dissolved in 0.1 M HCI, and that the specific heat of
0.1 M HCl is 0.99894 cal/g - deg at 25°C.
B. Real reaction measurements with the Parr calorimeter:
21
Follow steps 2-17 in the procedure above for each salt, using 100g of water for the
solvent; make a trial run for each salt to estimate the calorimeter span. Weigh
samples of each salt to give final concentrations as shown in the table below:
Final Suggested Span Setting
Concentration For Trial Run
Ba(NO3)2 0.02 M Lc
NaySOg 01M 0.25°C
NaNO; 0.04 M re
For runs involving a mixture of Ba(NO3). + Na,SO,, weigh samples to give the
same final concentration of Na” and NO, ions as above; use 1°C as the span setting
for the trial run.
Note: Ideally, you would want the same final concentration of NazSO4 as Ba(NO3)2
when determining the heats of solution of the pure salts so that the solutions have the
same ionic strength. However, the small amount of heat evolved when NazSQq is
dissolved makes this impractical; any error introduced here should be very small.
Remember, at the end of the day: Carefully wash and dry all wetted parts of the cell
assembly and the Dewar flask. Place the Dewar in the calorimeter can; replace the
cover and drive belt. Check that the recorder pen is lifted from the chart, replace the
pen cover, and turn off the calorimeter and recorder.
Determine AT, as outlined in steps 1-4 of the calculation section; the energy change
is calculated as above:
Q= AT, - Coat + OMg1y (Ceotv »
and the molar enthalpy change at the reaction temperature is calculated by
an =22
n
where n is the number of moles.
22