Docsity
Docsity

Prepare for your exams
Prepare for your exams

Study with the several resources on Docsity


Earn points to download
Earn points to download

Earn points by helping other students or get them with a premium plan


Guidelines and tips
Guidelines and tips

Lab Writeup Instructions - Optical Spectroscopy | PHYS 552, Lab Reports of Optics

Material Type: Lab; Class: Optical Spectroscopy; Subject: Physics; University: University of Illinois - Urbana-Champaign; Term: Unknown 1989;

Typology: Lab Reports

Pre 2010

Uploaded on 03/16/2009

koofers-user-nd9
koofers-user-nd9 🇺🇸

10 documents

1 / 15

Toggle sidebar

Related documents


Partial preview of the text

Download Lab Writeup Instructions - Optical Spectroscopy | PHYS 552 and more Lab Reports Optics in PDF only on Docsity! Physics 552 Optical Spectroscopy (Fall 08) - 1 - Lab Writeup Instructions Due at the beginning of your next lab section. Labs will generally be due the next week at your section. • You can either write on this revised lab writeup directly, or on a separate sheet as long as questions are clearly numbered. • Tables and plots can be printed out and attached at the end of the lab report. • You only have to answer one of either Q3 or Q7, depending on which section of the lab you did first. • Your data will be downloadable via the Labs section of the class webpage. http://online.physics.uiuc.edu/courses/phys552/fall08/ In general, for the Beckman DU experiment, if you did not get “good” results from your data, you should still analyze/present your own data, give reasonable conjectures as to why your data could be off, and explain what you would do differently to test your conjecture. (Note, that the whole method is too crude is not an acceptable reason. I was able to measure concentrations within 5% for solutions A and B using the Beckman.) If you want, you can feel free to analyze and present another group’s data as long as 1) you’ve analyzed and presented your own data and 2) you clearly acknowledge where the data came from. As for the monochromator concentration measurements, don’t worry too much about the actual concentrations you measured. Just clearly present your data. The graduated ND filter calibration is presented at the end. The optical densities (from which you can calculate concentrations) of the fluorescein solutions are in the table below: Finally, the first thing you need to do as part of your lab writeup assignment is to set up your Active Directory password. Go to the following site and follow the instructions: http://www.ad.uiuc.edu/ Name OD at 490 nm 1 1.85 2 0.525 3 0.245 4 0.957 Physics 552 Optical Spectroscopy (Fall 08) - 2 - Lab 1 – Monochromators and such… I. Introduction This lab is exploratory and introductory. If you use absorption spectrometers and fluorometers in your everyday research, here’s your chance to get in under the hood. If you don’t, now’s your chance to get acquainted with these instruments on a rudimentary level. You should approach this lab with an exploratory, hands-on perspective. We will be doing some approximate absorption measurements manually, without the automation of modern instruments Topics Covered • Monochromators • Wavelength Dispersion Elements o Diffraction Gratings • Fluorescein – absorption/fluorescence References 1) Jeremy M. Lerner, “Imaging spectrometer fundamentals for researchers in the biosciences - a tutorial” http://www.lightforminc.com/ImagingSpectrometerFundamentals.pdf Accepted for publication in the journal "Cytometry" (http://www3.interscience.wiley.com/cgi-bin/abstract/112593104/ABSTRACT) 2) The Instrument Project: UV-visible spectroscopy. http://www.wooster.edu/chemistry/is/brubaker/uv/default.html 3) The useful information from a leading monochromator manufacturer http://www.newport.com/Monochromators-and-Spectrographs/370636/1033/catalog.aspx 4) Eugene Hecht, Optics II. Monochromator Experiment In this part of the lab you will be playing with a Bausch & Lomb grating monochromator (circa 1950s). First, familiarize yourself with the operation of the monochromator. Then you will use the monochromator to explore the qualitative absorption properties of a fluorescein, a highly absorbing and fluorescent compound. Finally, you will perform a rudimentary absorption measurement by comparing the intensity of the transmitted light to a known reference by eye. • While these monochromators are old, the mirrors and diffraction gratings are valuable! Do not touch/clean the mirror or grating surfaces. Gratings are destroyed and mirror surfaces seriously damaged by fingerprints. They’ve survived this long, so give ‘em another 60 yrs. • The filters used in this lab are Prof. Clegg’s research equipment. They are expensive! Do not touch Physics 552 Optical Spectroscopy (Fall 08) - 5 - From Ref 1. λω ka =)sin(2 Note that this is for a special case where the incident angle equals the diffraction angle, the Littrow configuration. If you want, look for the second order maxima. No extra points will be given for your discovery so you’ll want to do this after you’re done with Lab 1. What if you wanted to use a grating monochromator to take a spectrum from 100 nm to 500 nm (e.g., the absorption spectrum of fluorescein, which you will be doing later in the lab). Say you scan to the first order maximum of 400 nm. a) Is that the only wavelength that would come out of the exit slit of the monochromator? Would you see other wavelengths? What order maxima would they be? b) How could you eliminate other wavelengths if they are present? (Hint: what kind of optical equipment could you use?) Q2) A. Using the Monochromator for Absorption Measurements In this section, we use the monochromator to make some rough, rudimentary absorption measurements. We will use fluorescein, a highly absorbing compound. The energy from any incident light is transferred to the molecule, sending it into an excited state. We get the absorption by measuring the amount of transmitted light that is not absorbed (light that is scattered would also not be transmitted). In lab 2, we will go into more detail regarding the theory behind absorption. For now, let’s just get down some working “facts”. Define the optical density of a solution by the log of the ratio of the incident and transmitted intensities )log( 0 IIOD = ODII −⋅= 100 The optical density is composed of absorbing and scattering components SAOD += Physics 552 Optical Spectroscopy (Fall 08) - 6 - For this lab, we assume scattering is negligible, i.e., OD ~ A. Intuitively, you would expect the absorption to depend on the number of absorbing molecules the light beam encounters, and hence on the concentration of absorbers (C, mol/L), path length through the sample (x, cm). So xCA ⋅⋅= ε where ε, (L/mol-cm), the molar extinction coefficient, is a fundamental property for each compound. This is known as the Beer-Lambert Law. For now, A~OD. ε(λ): extinctioncoefficient concentrationC: incident intensity I0 (λ) transmitted intensity I (λ) x: path length sample For fluorescein at 490 nm, ε =93000 M-1cm-1. x = 1 cm for the cuvettes we are using. 1) Crude Absorption Spectrum N.B. If you started off with the Beckman DU and already drew a guestimated absorption spectrum, you can skip this part. Although, it is still a good idea to quickly go through the motions to get a feel for the instrument. Scan the monochromator through the visible spectrum to get an idea of the intensity of the lamp at each wavelength. The monochromator has 235 nm offset (that is if you want to select 500 nm you’ll need to adjust the knob to 735 nm). Using the fluorescein solution in the cuvette marked CONC (concentrated), again scan the monochromator through the visible spectrum. It might help to get an idea how much light gets absorbed at each wavelength if you cover the top half of the exit slit with the cuvette and then look at the two halves at the same time. See the next section if you have a hard time setting this up. Sketch a very rough absorption spectrum for fluorescein (OD vs. λ) (Main feature to get: any absorption peaks. I know this will be very rough. Just roughly use red 700 nm, green 550 nm and violet 400 nm as landmarks.) Q3) Rough Estimate Absorption Spectrum of Fluorescein (OD vs. λ) You should give rough numbers for the λ axis. You don’t have to do that for the OD axis. You will measure this more quantitatively later with the Beckman DU and a CCD camera. 2) Concentration Measurements One of the uses of absorption measurements is determining solution concentrations. Here you will do this by eye, matching the intensity transmitted by the fluorescein solution to that transmitted by a graduated neutral density filter. Neutral density means it transmits (roughly) the same amount at all wavelengths. Graduated means there is a spatial gradient in the OD across the filter. The graduated ND filter you are given is linear, so given the distance d along the filter, you know OD (d). The procedure is as follows: Physics 552 Optical Spectroscopy (Fall 08) - 7 - • The absorption peak of fluorescein is at 490nm. • Place the cuvette with fluorescein marked A into the clamp on the ring stand. Adjust the height of the cuvette so that it covers the top half of the exit slit. Place the graduated ND filter A on the lab jack in front of the exit slit and adjust it’s height until it covers the bottom half of the exit slit. • Slide the filter left and right until you find d for which the transmitted intensity matches that of the cuvette. Record d, as well as the uncertainty (cm) in your measurement (e.g., d = 5 ± 0.5 cm) . • Using the conversion table/equation provided (at end), convert your measured d into an OD. Then calculate the concentration of the fluorescein solution. • Hard to judge, isn’t it? Everyone in your group should make an independent measurement. There are two samples (A and B). If you have a large group, have different people measure different samples, but make sure each sample has more than one independent measurement so you can take the mean and std dev. Q4) a) For your own measurement, calculate the OD, and convert the uncertainty in d to an uncertainty in OD. b) What concentration does your measured OD correspond to? c) Record all measured d for your group for each sample measured. Calculate the mean OD and standard deviation. Compare this value to the actual concentration given on page 1. d) Do a propagation of error calculation. i) If you had a 10% error in measuring d, what is the error in your calculated concentration? ii) With this experiment setup, it comes out that you basically write down an OD directly, although you are really making a comparative judgment of the intensity of the transmitted light. What if you were measuring an intensity directly (i.e., you were counting photons) with a photodetector. In that case, if you had a 10% error in I, what error does that translate to in your measured concentration? Assume Io is known exactly. Physics 552 Optical Spectroscopy (Fall 08) - 10 - B. Quantitative Absorption Measurements If your group started with the Beckman first instead of the monochromator, read the intro to section II. B. for the necessary absorption measurement equations. 1) Crude Absorption Spectrum N.B. If you started off with the monochromator and already drew a guesstimated absorption spectrum, you can skip this part. Although, it is still a good idea to quickly go through the motions to get a feel for the instrument. Scan the Beckman through the visible spectrum to get an idea of the intensity of the lamp at each wavelength. Using the fluorescein solution in the cuvette marked CONC (concentrated), again scan the monochromator through the visible spectrum. Sketch a very rough absorption spectrum for fluorescein (OD vs. λ) (Main feature to get: any absorption peaks. I know this will be very rough.) Q7) Rough Estimate Absorption Spectrum of Fluorescein (OD vs. λ) You will measure this more quantitatively later with a CCD camera. 2) Concentration Measurements We will be using the CCD camera to measure the intensity of the transmitted light. The procedure will be as follows: • Use PixeLink Capture to save a TIFF image of the transmitted light (see Appendix B for software instructions). • Measure a mean intensity of the transmitted light from the saved TIFF file using the image analysis program NIH ImageJ (which is free to download (http://rsb.info.nih.gov/ij/) - if time becomes tight, you may need to analyze the images later with this program on your own computer). Ask your TA to walk you through the use of the programs if you have any questions or problems. When taking images with Capture, you should make sure the images are not overexposed (that is the recorded image is not overly bright and saturated). Do this by changing the exposure time rather than the gain. Again ask your TA. • Take a picture (and save it on your computer) for each of the four “samples”: air (no cuvette), empty cuvette, water, fluorescein cuvette 4. There is a slideable cuvette holder that you can quickly change between samples, but you may need to swap in some cuvettes. • Use ImageJ to record a mean intensity of the transmitted light for each of the five samples. Save the resulting measurement table. Physics 552 Optical Spectroscopy (Fall 08) - 11 - Q8) a) • Present your data. o TIFF files should be available for your TA. Let him/her know where you’ve saved the files at the end of your time with the Beckman DU so they can be transferred over to the class server for future reference. o The measurement table from your intensity analysis in ImageJ, with each measurement clearly labeled. • Do the relative transmitted intensities for each sample make sense compared to the others? The empty cuvette shows a measured OD. Is this absorption or something else? How does it compare to the cuvette of water? b) Calculate the concentration of your fluorescein sample and compare with the given values. Use the intensity measured for water as your Io. Why should it make sense to “subtract out” the water? 3) Quantitative Absorption Spectra (Instructions and Questions) • Pick 5 or 10 (as time allows) roughly equally spaced points in the visible spectrum and measure the intensity for a fluorescein solution using the CCD. Make the same measurements for a water solution. If time is tight, ask your TA for instructions o Be sure to include 490 nm and 514 nm. • Extract the mean intensities from your snapshots using ImageJ as you did in the last section. Have the TA put your TIFF files on the server, and include the table of measured intensities from ImageJ. Physics 552 Optical Spectroscopy (Fall 08) - 12 - Q9) a) Present your table of measured mean intensities, including any exposure time data and the normalized intensities. Exposure times were changed, so remember to divide the Mean by the exposure time for each measurement. b) Plot the quantitative, normalized absorption spectrum. • Calculate the OD for each wavelength measured. Use the water measurement at each wavelength as your Io. Make sure you normalize for any different exposure times. • Compare this to the actual spectrum for fluorescein F1300. Download the this spectrum in on the course web site http://online.physics.uiuc.edu/courses/phys552/fall08/Labs • Place both your measured spectrum and the downloaded spectrum on the same plot, print out and attach to your lab writeup. c) Given that you understand the point of the mystery light question, why don’t you see a peak at 514 nm here? The Beckman DU ushered science into a George Jetson era. [From the Beckman Model DU Spectrophotometer Instruction Manual (305-A), Beckman Instruments Inc., (Fullerton, CA).] IV. Synthesis Questions You should look over these questions and think about them before you leave lab, so you can check over things that you might need.
Docsity logo



Copyright © 2024 Ladybird Srl - Via Leonardo da Vinci 16, 10126, Torino, Italy - VAT 10816460017 - All rights reserved